FLUKE SAFETY RECALL
Through quality assurance tests, Fluke has discovered a potential short
circuit connection on the circuit board. This may lead to inaccurate voltage
readings including a low or no voltage reading on a circuit energized with a
hazardous voltage. This finding could create a hazardous situation if the user
were to contact live voltage based on an erroneous instrument reading.
Required Actions:
1. Stop using your affected 33X series clamp meter(s) even if you have not
experienced problems.
2. Verify your clamp meter is affected using the serial numbers below.
Fluke-337 (All models including 337A)
Serial# 96070000-98890607
Fluke-336 (All models including 336A)
Serial# 96220000-98900400
Fluke-335 (All models including 335A)
Serial# 97010000-98860104
Fluke-334 (All models including 334A)
Serial# 97010000-98870101
Fluke-333 (All models including 333A)
Serial# 97010000-98880240
3. Go to:
FLUKE Safety Recall and follow the directions provided.
For more information, call (888) 98-FLUKE (35853) 7 a.m. to 4 p.m. Pacific Time,
Monday-Friday.